Table 1.2. Components and controls of the Analyzer (cont.)
K
L

Specifications

The following are specifications for the Analyzer. Please contact your Fluke
Biomedical service representative for more information regarding the device

specifications.

Line Voltage
Range .....................................................................90 – 240 V ac 50/60 Hz
Accuracy................................................................. ±3 % of reading, ±1 LSD
Load Current
Range .....................................................................0 – 19.99 A
Accuracy.................................................................±5 % of reading, ±1 LSD
1-10
800.561.8187
Provides a means for inputting the chassis
cable with its clip for connection to the DUT
chassis or enclosure. The chassis ground
Chassis
resistance is measured with the FUNCTION
Connector
switch in the RESISTANCE position, and the
chassis leakage current is measured in the
CHASSIS position.
Used to make point-to-point measurements
with optional leads available for purchase. For
leakage and voltage measurements, the black
cable with the clamp with black insulation is
attached to the CHASSIS connector, and the
Dual Connector
black cable with the clamp with red insulation to
the DUAL connector. For point-to-point
resistance measurements, the black cables
with clamps with black insulation are attached
to the DUAL and CHASSIS connectors.