Updating Attached Reader Firmware

Important! All settings may be reset to manufacturer default values during the update process,
so it is recommended to take a note of the present settings prior to updating the RM310 or IS1001
readers.

Updating Attached RM310 Reader Firmware

New or updated firmware may periodically become available for the RM310 reader. To update
the firmware:
1. Start the BioTerm communication program.
2. Make a connection to the BLE Data Logger using RS-232 communication port (RM310
firmware update through Bluetooth connection is not supported).
3. Enter the LPM command and type Y after confirmation question appears to switch the logger
into RM310 programming mode.
4. Close the BioTerm communication program.
5. Use Allflex Reader Programmer software to update RM310 reader firmware (refer to RM310
User Guide for details).
6. After the update is complete reset power to the BLE Data Logger.
Note: To exit RM310 programming mode power to the logger must be reset. The logger will exit
RM310 programming mode automatically in 10 minutes.

Updating Attached IS1001 Reader Firmware

New or updated firmware may periodically become available for the IS1001 reader. To update
the firmware:
1. Start the BioTerm communication program.
2. Make a connection to the BLE Data Logger using either RS-232 communication port or
Bluetooth.
3. Enter the CDC command to configure the connection as the reader communication port.
4. Select Tools > Update Device Firmware.
5. Go to the location of the new firmware file, select it and then click on the Open button.
Make sure to use the appropriate file for your IS1001 model:
IS1001: This model uses IS1001 Application Firmware v1.x.x.
IS1001-12V: This model uses IS1001 Application Firmware v2.x.x.
IS1001 BLE Data Logger User Manual
Updating Attached Reader Firmware
16