Guidance Tables

Recommended Separation Distances between
Portable and Mobile RF Communications Equipment and the Prima GN for
ME Equipment and ME Systems that are NOT Life-supporting.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Portable and Mobile RF Communications Equipment and the Prima GN
The Prima GN is intended for use in the electromagnetic environment in which radiated RF
disturbances are controlled. The customer or user of the
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF Communications Equipment (transmitters) and the Prima GN as recommended below,
according to the maximum output power of the communications equipment.
Max Output Power
of Transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d)
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (w) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
6165000-795
(continued)
Recommended Separation Distances for between
Separation (m)
150kHz to 80 MHz
d=(3.5/v1)(Sqrt P)
0.1166
0.3689
1.1666
3.6893
11.6666
Prima GN can help prevent
Separation (m)
80 to 800 MHz
d=(3.5/E1)(Sqrt P)
0.1166
0.3689
1.1666
3.6893
11.6666
PRIMA GN
Prima GN
Separation (m)
800MHz to 2.5GHz
d=(7/E1)(Sqrt P)
0.2333
0.7378
2.3333
7.3786
23.3333
Issue 1.4
Printed on March, 2016