Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Prima GN is intended for use in electromagnetic environment specified below. The
customer or user of the Prima GN should ensure that it is used in such an environment.
Immunity
Test
ESD
IEC 61000-4-2
EFT
IEC 61000-4-4
Surge
IEC 61000-4-5
voltage
Dips/Dropout
IEC 61000-4-11
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
6165000-795
(continued)
Guidance and Manufacturer's Declaration
Electromagnetic Immunity
All Equipment and Systems
IEC 60601
Compliance
Level
Test Level
±6kv Contact
±6kv Contact
±8kv Air
±8kv Air
±2kv Mains
±2kv Mains
±1kv I/Os
±1kv I/Os
±1kv Differential
±1kv Differential
±2kv Common
±2kv Common
>95% Dip
>95% Dip
for 0.5 Cycle
for 0.5 Cycle
60% Dip
60% Dip
for 5 Cycles
for 5 Cycles
30% Dip
30% Dip
for 25 Cycles
for 25 Cycles
>95% Dip
>95% Dip
for 5 Seconds
for 5 Seconds
3A/m
3A/m
Electromagnetic
Environment
Floors should be wood, concrete or ceramic
tile. If floors are synthetic, the R/H should be at
least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the Prima GN requires continued operation
during power mains interruptions, it is recom-
mended that the Prima GN be powered from
an uninterruptible power supply or battery.
Power frequency magnetic fields should be that
of a typical commercial or hospital environment.
PRIMA GN
Prima GN
- Guidance
29
Issue 1.4
Printed on March, 2016