APPENDIX 3
RS232 Port Specifications
Baud Rate:
Parity :
Bits:
Stop Bit:
Data Format:
*ALL X's are numerals from 0-9.
X X__ (space) _ X.XX (space) X.XX (space) __ XX.X(space)
repeats
The first field (XX) indicates temperature in
supplied.
The second field (X.XX) indicates heater time, hours.minutes. The display "C.CC" denotes a
continuous heat run. The display "E.nd" denotes the end of the heat run
The third field (X.XX) indicates run time, hours.minutes. The display "E.nd" denotes the end
of the run.
The fourth field indicates vacuum in torr units. If the vacuum is less than 1 torr, the format is
.XXX. If it is greater than 1 torr, the format is XX.X
For example:
5 6 __ (space) _ 3.30 (space) __3.59 ________ (space)
repeats)
This represents a temperature of 56
of 3 hours and 59 minutes and vacuum level of 15.8 torr.
9600
None
8
1
C, a heater time of 3 hours and 30 minutes, a run time
°
C. The display "no" denotes no heat being
°
15.8 (space)
sequence
(sequence