SmartLite™ PS
4.
Adjust the LED Tip: The LED Tip rotates by 360°, thus the position of light emission
may be freely adjusted. The tip must be adjusted to an angle that most optimal curing of
the restoration is possible. The tip must not touch the restoration material but should be
positioned as close as possible to the restoration.
Polymerisation: Shortly press the ON/OFF key [1.1]. The start of the 10-second
5.
exposure cycle is indicated by an acoustic signal. The end of 10-second exposure cycle
is likewise indicated by an acoustic signal.
For detailed information on curing times for DENTSPLY light cured products, please see
the Curing Guide supplied. For products other than those from DENTSPLY, please refer
to the curing times given by the respective manufacturer.
Wait for at least 30 seconds between exposures. If time intervals between exposures are
too short, this may lead to a potentially tissue injurious overheating of the LED Tip.
10.3

Maintenance

Prevention of cross infection, cleaning, disinfection, and storage
The use of the sleeve is an additional precautionary measure against contamination and
does not substitute disinfection of the device.
After use, remove sleeve. Clean and disinfect LED Tip and hand piece with commercial
alcohol based surface disinfecting solution (Dürr FD 322, Henkel Incidin Liquid, S&M
Mikrozid Liquid, B.Braun Meliseptol rapid). Keep other solvents or flammable liquids as well
as intense sources of heat away from the unit as they may damage its plastic housing.
Always safeguard charger, hand piece, and LED Tip against moisture as this may cause
electrical short-circuit or dangerous malfunction.
Light output control
Make sure that the LED aperture is clean and scratch-free, otherwise light output will be
reduced and may be insufficient for proper curing of the material.
Control of light efficiency is recommended (e. g. by using CureRite™). The CureRite output
should exceed 730 mW/cm².
Battery maintenance
The battery needs a total of 10 hours to be fully re-charged.
When the battery is charging, the battery light flashes green.
Upon complete recharging, the battery light remains continuously green.
Scientific Compendium
20 / 22