Chapter 4
Instructions for Use
4-8
T3/G3
this reason, the centrifuge is equipped with an imbalance detector.
In case of excessive imbalance, the load imbalance indicator LED is
illuminated and the brake is applied immediately. The rotor will be
decelerated to rest in few seconds.
As soon as the motor stops, open the centrifuge and redistribute the
samples to produce an equal weight on diametrically opposite sides.
Also remember that the pins must be softly lubricated. Inadaquate
lubrication of the pins may cause incorrect swinging of the buckets
resulting in imbalance phenomen. See chapter 6.6 for details.
If the message persists, despite your efforts to balance the load, call
your Thermo Service representative.
Thermo Fisher Scientific