40 Troubleshooting

System problems

The electrocardiograph won't turn on when it is plugged in
Causes
The AC power connection is faulty.
An AC fuse is blown.
There is no AC power.
Actions
Check the AC power source.
Check the AC fuses.
The electrocardiograph won't turn on when it is unplugged
Causes
The battery is disconnected or incorrectly connected.
The battery is low, not charging, depleted, or bad.
Actions
Check the battery connections.
Recharge the battery.
Replace the battery.
The electrocardiograph shuts down during printing
Causes
The battery is low or bad.
Actions
Recharge the battery.
Replace the battery.
The electrocardiograph prints fewer than 10 reports on a full battery
charge
Causes
The battery is degraded.
Actions
Replace the battery.