Welch Allyn CP 50 Directions For Use Manual page 43

12-lead resting electrocardiograph
Hide thumbs Also See for CP 50:
Table of Contents

Advertisement

Directions for use
Lead alert or square wave
A dot might be flashing on the lead-status screen. Or one or more leads might appear as
a square wave.
Causes
Electrode contact might be poor.
A lead might be loose.
A lead might be defective.
Actions
Replace the electrode.
Verify that the patient's skin has been properly prepared.
Verify that electrodes have been properly stored and handled.
Replace the patient cable.
Troubleshooting 39