Directions for Use
Diagnostic Desk System
Intended Use
This product is intended for use in outpatient care, emergency
rooms, observation rooms, hospital wards, and community clinics
to measure patients' body temperature and blood pressure and
fundus and ear canal examination.
Symbol Descriptions
AC input power
DC input power
DC output power
Recycle used batteries properly and in accordance
with local regulations.
Do not dispose of batteries in refuse containers.
China's directive (RoHS) markings for control of
pollution caused by electronic information
products.10 indicates the maximum number of years
of environmentally friendly use.
TYPE BF – Indicates this is a product with Type BF
application components.
Medical equipment that conforms with UL60601-1
specifications for electric shock, fire, and mechanical
hazards.
The equipment is not waterproof.
Class II
Date of manufacture
Batch code
Catalogue number
Warnings and Cautions
WARNING: Read and understand the instructions
before using this product. Failure to understand the
operating requirements for this product can result in
injury to you or the patient and can damage the
instrument.
WARNING: Explosion hazard. Do not operate this
product in the presence of flammable anesthetics
or other substances that are flammable in
combination with air, oxygen-enriched
environments, or nitrous oxide.
WARNING: This is a class B product. In a medical
facility this product can cause radio interference, in
which case the user might need to take corrective
measures.
WARNING: Do not modify this equipment. Any
modification of this equipment can lead to patient
injury. Any modification of this equipment voids the
product warranty.
WARNING: Electric shock hazard. Do not
autoclave the device or accessories. The device
and the accessories are not heat-resistant.
WARNING: When the device is powered on, do not
simultaneously touch the contact pins and the patient.
CAUTION: Only authorized Welch Allyn Service
Centers can perform service on this device.
CAUTION: To make sure that the device can be
unplugged from the AC power source, never block
access to the wall socket.
CAUTION: Refer to manufacturer's directions for
use for the diagnostic equipment.
Bare hand operation prohibited
Environment
Operating conditions
Temperature Range
Transport/Storage
Temperature Range
Classification
:
Safety Class II
Internally powered equipment
Type BF application components
Protection IPX0
None AP/APG equipment
Continuous Operation
Electrical Power Supply
100-240/V, 50-60 Hz, 0.18A Max
5V, 1A
Charger:
5V, 1A
5V, 0.5A
Humidity Range
Atmospheric
Pressure Range
Humidity Range
Atmospheric
Pressure Range
31