44 Specifications
Item
Report formats, external
printer*
Auto ECGs
Average cycles
ECG storage (in test
directory)
Patient storage (in
worklist, CP 50 Plus only)
Frequency range
Digital sampling rate
Pacemaker detection
Power requirement
AC fuses
Rechargeable battery
Filters
High-performance
baseline
Muscle tremor
AC interference
Safety, EMC, and
regulatory compliance
Standard connectivity
CP 50™ and CP 50 Plus™ 12-lead resting electrocardiograph
Specification
2x6 + 1R @ 50 mm/s, No print
3x4 @ 25 mm/s, 3x4 @ 50 mm/s, 3x4 + 1R @ 25 mm/s, 3x4 + 3R @ 25 mm/s,
3x4 - 5.0s @ 25 mm/s, 3x4 - 5.0s @ 50 mm/s, 6x2 - 5.0s @ 25 mm/s,
12x1 - 5.0s @ 25 mm/s, No print
3x4 + 3R @ 25 mm/s, 3x4 + 3R @ 50 mm/s, 2x6 + 1R @ 25 mm/s,
2x6 + 1R @ 50 mm/s, 3x4 + 2R @ 25 mm/s, 3x4 + 2R @ 50 mm/s,
6x2 + 2R @ 50 mm/s, No print
At least 25 ECG tests
Up to 50 patients
0.3 to 150 Hz
> 1,000 samples/second/channel
ANSI/AAMI EC11
Universal AC power supply 110-240 V~, 50/60 Hz ~, 1.5 A maximum
Time-lag type, 2.0-amp 250-V rating, Littlefuse 0218002P or equivalent
10.8V, 1.9Ah minimum, 3-cells Li-ion "smart" battery. A fully charged battery prints up to 25
ECGs with one extra copy per test. Recharge time to 90 percent capacity is 6 hours.
0.5 Hz
35 Hz
50 Hz or 60 Hz
ANSI/AAMI EC11**
CAN/CSA C22.2 No. 601.1
CAN/CSA C22.2 No. 601.1.1
CAN/CSA C22.2 No. 601.1.2
CAN/CSA C22.2 No. 601.1.4
CAN/CSA C22.2 No. 601.2.25
Mini USB client and USB host, Ethernet (optional - available with CP 50 Plus only)
UL60601-1
IEC/EN 60601-1
IEC/EN 60601-1-1
IEC/EN 60601-1-2
IEC/EN 60601-1-4
IEC/EN 60601-1-6
IEC/EN 60601-2-25 ***
IEC/EN 60601-2-51 (2x6 lead arrangement)