20 Setup
To connect to a workstation via USB cable
Connect the electrocardiograph to any of the workstation's USB ports. No software
configuration is required. The electrocardiograph is ready to communicate.
To connect to a network-connected workstation via Ethernet cable
1. Connect the electrocardiograph to the same network as the workstation. If you need
help, consult with your network administrator.
2. Go to the CP 50 connectivity screen
a. Touch the menu button.
b. Touch the Settings > System > Network > Device.
3. Fill in the screen.
4. Touch the Test tab
5. Connect to the network.
a. In the Host Address field, enter the host IP address.
b. Touch Ping Test. The electrocardiograph pings the host.
6. Touch the Server tab.
7. Choose one of these to specify how to get server IP information:
8. Touch Server Test.
The electrocardiograph connects to the server.
WARNING To reduce the risk of fire and shock hazards,
connect Ethernet cables within the boundaries of a single
building. Ethernet cables that span multiple buildings can
introduce fire or shock hazards unless fiber optic cables or
lightning arrestors are properly installed or other appropriate
safety measures are taken.
If DHCP is enabled for the network, select the DHCP checkbox.
If DHCP is not enabled, fill in the fields: static IP address, subnet mask, and
default gateway.
Obtain server IP information automatically: Select this option if you want the
device to get the IP address, or if you don't know the server's IP address.
Server IP address: Fill in this field if your server IP address is available.
CP 50™ and CP 50 Plus™ 12-lead resting electrocardiograph