Inspecting The Equipment; Testing The Electrocardiograph; Replacing The Battery - Welch Allyn CP 50 Directions For Use Manual

12-lead resting electrocardiograph
Hide thumbs Also See for CP 50:
Table of Contents

Advertisement

34 Maintenance
3.
4.

Inspecting the equipment

Perform the following inspections daily.
Check for cracked or broken patient cable, patient electrodes, power cord,
communications cables, display, and enclosure.
Check for bent or missing pins on all cables.
Check all cable and cord connections; reseat if any connectors are loose.

Testing the electrocardiograph

Welch Allyn recommends verifying proper operation of the electrocardiograph once a
year to ensure reliability. See Verifying proper operation.
Whenever the electrocardiograph is serviced or whenever problems are suspected,
verify continued electrical safety of the device using IEC 60601-1 or ANSI/AAMI ES1
methods and limits.
Test for the following:
Patient leakage current
Chassis leakage current
Earth leakage current
Dielectric strength (AC and patient circuits)

Replacing the battery

Replace the battery under these circumstances:
It loses its charge quickly.
You have charged it, and the electrocardiograph still does not turn on when
unplugged.
The first time that you press the power button after installing a new battery, the
electrocardiograph goes through some diagnostic tests that cause it to take longer than
usual to power up.
Discard the old battery appropriately. Contact your local authorities concerning recycling.
Dampen a cloth with any of the acceptable cleaning solutions, and wipe the exterior
of the patient cable, reusable electrodes, and electrocardiograph. Dry all
components with a clean, soft cloth, or paper towel.
Before you turn on the electrocardiograph again, wait at least 10 minutes for all
traces of liquid to evaporate.
WARNING Only qualified service personnel should perform leakage
current tests.
CP 50™ and CP 50 Plus™ 12-lead resting electrocardiograph
WARNING The patient cable and reusable elctrodes should be
cleaned between each patient.