Keysight M8290A Getting Started Guide
External local oscillator input and output (Option 800)
LO input wavelength range
LO input power range
LO output power
1
128 samples are unavailable, resulting in 511872 samples per channel effectively
available.
2
Valid at EVM = 32.5% for 32 GBaud DP-QPSK corresponding to raw BER = 1E-3,
boost mode off.
3
At maximum laser power.
4
Adjusted baseband frequency range available for signal analysis.
5
Valid at the following reference conditions
•
Sampling rate 92 GSa/s
•
Optical continuous wave signal at optical input port
•
Signal power > 0 dBm
•
Optical frequency is offset by 2.5 GHz from local oscillator frequency
•
Vector analyzer I-Q spectrum span set to 12.5 GHz
•
QPSK demodulation
•
10 Gbaud symbol rate
•
PolStokesAlign set to "Single Polarization"
•
KFPhaseTrack with carrier phase variance set to 1E-4
•
Result length set to 500 symbols
•
Raised cosine filter selected as reference filter
•
25° C ±5 K environmental temperature
6
Valid at reference conditions as stated above, except for:
•
Optical frequency is offset by 10 GHz from local oscillator frequency
•
Vector analyzer I-Q spectrum span set to 50 GHz
•
40 Gbaud symbol rate
Table 5
M8292A Modular Optical Modulation Analyzer specifications (Continued)
Trigger input
Input range
Threshold
Range
Resolution
Specifications
1527.60 to 1570.01 nm
–3 dBm to +17 dBm
> +9 dBm
Typical values
–4 V to +4 V
–4 V to +4 V
10 mV
100 mV
6
51