5 Sensor setup

IMPORTANT (NOTE)
Perform this procedure on existing sensor(s) only.
Sensors are setup / configured individually. If installing a
new / replacement sensor, refer to Section 4, page 7.
1.
Connect the ATS430 sensor to the transmitter's EZLink
connector – see transmitter Operating instruction
OI/AWT440-EN.
2.
At the AWT440 transmitter, press the
Operator Page menu, then select Enter Configuration to
display the Access Level page.
Use the
key to select the Advanced menu item and
press the
key (below the Select prompt).
If the Sensor Setup menu is not displayed use the
keys to scroll to it:
Menu
Sensor Setup
Exit
Press the
key (below the Select prompt).
8 OI/ATS430–EN Rev. H | Aztec ATS430 | Turbidity sensor
key to display the
Select
3.
The Sensor Setup page is displayed:
Sensor Setup
Sx: Turb TSS
Back
Ensure Sx :Turb TSS is highlighted and press the
(below the Select prompt).
The Sx :Turb TSS: Turbidity menu page is displayed:
Sx :TURB TSS
Tag
PV Type
Units
/
Range High
Range Low
Clean Interval
Back
4.
Proceed with sensor setup – see Section 6.1, page 10
for parameter options.
3
2
1
Select
3
2
1
Select