6.2.2 Initiating the verification at the transmitter

1.
At the Calibrate level, press the
prompt):
Menu
Calibrate
Exit
The sensor selector menu is displayed:
Calibrate
S1:Turb TSS
Hold Outputs
Back
Highlight the sensor to be verified (for example
S1:Turb TSS) and press the
prompt).
2.
The menu options for S1:Turb TSS are displayed:
S1:Turb TSS
Sensor Verification
Turbidity Cal
Suspended Solids Cal
Restore Defaults
Back
3.
Use the
/
keys to select Sensor Verification
and press the
key (below the Select prompt).
The Sensor Verification screen is displayed:
Sensor Verification
Verification Value
4000 NTU
Next
4.
Press the
key (below the Edit prompt) to enter the value
of the verification puck.
12 OI/ATS430–EN Rev. H | Aztec ATS430 | Turbidity sensor
key (below the Select
Select
Select
key (below the Select
Select
Edit
Verification Value
04000
Max
4000
Min
0.000
Next
5.
Enter the turbidity value printed on the puck label and press
the
key (below the OK prompt).
The Start Verification screen is displayed:
Verification
Start Verification?
1000 NTU
Abort
Ensure the sensor is inserted in the verification tool
(see Section 6.2.1, page 11) and press the
(below the Continue prompt) to start the verification
routine. (To Abort verification, press the
6.
The Verification progress screen is displayed:
Verification
PV
1001 NTU
STD
1000 NTU
Abort
If the verification process completes successfully, a
verification successful message (Procedure Pass) is
displayed:
Verification
PV
1001 NTU
STD
1000 NTU
Procedure Pass
Exit
3
2
1
NTU
OK
Continue
key
key).