ABB Aztec ATS430 Operating Instructions Manual page 14

Turbidity sensor
Table of Contents

Advertisement

5.
Perform a sensor calibration at the transmitter:
— see Section 6.3.2, page 15 for a 1-point calibration
— see Section 6.3.3, page 16 for a 2-point calibration.
6.
When the calibration is complete, withdraw sensor / cap
assembly
H
from calibration pot
sensor
B
(a combined pull and twist action is the easiest
withdrawal method). Rinse all items with water and dry
thoroughly with tissue (not supplied).
Fig. 6.2 Calibration using the optional calibration pot
14 OI/ATS430–EN Rev. H | Aztec ATS430 | Turbidity sensor
G
. Remove cap
A
from
7.
Dispose of formazin solution safely in accordance with local
regulations.
200 ml (6.76 fl. oz)
formazin