3 Installation

3.1 Siting

IMPORTANT (NOTE)
— The sensor is supplied with a protective cover on the
end cap. The cover must be removed before the
sensor can be operational.
— When installing the sensor, ensure that the front
face of the sensor is submerged to at least 30 cm
(11.81 in.) and the sensor is at least 5 cm (1.96 in.)
away from any surface in all directions.
— When using extension cables, protect the
connections using heat shrink (for example,
HISA-18/6-PEX-CL or equivalent).
Fig. 3.1 Siting the sensor

3.2 Sensor dimensions

Dimensions in mm (in.).
10
(0.40)
40 (1.57)
Fig. 3.2 Sensor dimensions
Max.
210 m (64 ft.)
180
(7.09)
150
(5.90)
5 cm
(1.96 in.)
min.
5 cm
(1.96 in.) min.
10000 (393)
1 in. BSP thread
Aztec ATS430 | Turbidity sensor | OI/ATS430–EN Rev. H 5
Sample surface
30 cm
(11.81 in.)
min.
5 cm
(1.96 in.)
min.
1000 (39)
or
nominal