Page 1
It is the perfect starting point for pH or Redox measurements. Service Combining measurement, reference and temperature compensation elements in a single unit, the AP300 is an economical solution for a wide variety of industrial applications; including water, food, pharmaceutical and chemical.
1 Introduction 1.1 Purpose This instruction manual describes the installation and maintenance of the AP300 Series Industrial pH and Redox (ORP) Electrode Systems. 1.2 Sensors and Systems There are three main sensor types: — AP301 Standard twist-lock – insertion, flow-through, immersion —...
Dimensions in mm (in.) in. Dia. in. Hex. in. Hex. in. Hex. in. NPT 1 in. NPT 97.8 66.7 (3.85) (2.63) Hand Tight Wrench Tight Fig. 1.4 AP304/5 Compression Fitting Dimensions AP300 | Combination pH/Redox (ORP) sensors | IM/AP300 Rev. J...
Page 6
NPT Full Port Ball Valve in. Close Nipple AP305 Sensor Reducing Bushing Ball Valve Insertion Wrench Fitting – Installation Detail 28.6 Compression Fitting (1.13) Fig. 1.5 AP304/5 Sensors and Overall Dimensions IM/AP300 Rev. J | AP300 | Combination pH/Redox (ORP) sensors...
Fig. 2.1 Recommendations for Fouling Applications Note. — The flow of sample passing the sensor helps to keep the sensor clean. — Sensors should be positioned such that they are always immersed in the sample. AP300 | Combination pH/Redox (ORP) sensors | IM/AP300 Rev. J...
AP305 – optimize self-cleaning by using flat glass and notched body Sample flow through pipe Bulb Sample Flow Bulb Flat Glass Fig. 2.4 Installing the AP304/5 – Ball Valve and Compression Fitting IM/AP300 Rev. J | AP300 | Combination pH/Redox (ORP) sensors...
Page 11
4. Add together measurements from step 3 and, measuring from the tip of the sensor, mark off total onto the shaft of the sensor. Fig. 2.5 Installing the AP304/5 – Installation Measurements AP300 | Combination pH/Redox (ORP) sensors | IM/AP300 Rev. J...
Redox (ORP) without temperature compensation Blue Glass Electrode Male BNC Reference Electrode Black Blue Metal Electrode Redox Reference Electrode Black Fig. 3.1 Cable Terminations IM/AP300 Rev. J | AP300 | Combination pH/Redox (ORP) sensors...
Also ensure when moving from one buffer solution to the next to wash the electrodes and dry them carefully using a soft tissue. IM/AP300 Rev. J | AP300 | Combination pH/Redox (ORP) sensors...
If it is necessary to remove the electrode from the sample line, fill the retained protective cap with buffer solution and cotton wool, or equivalent, and fit it to the sensor. AP300 | Combination pH/Redox (ORP) sensors | IM/AP300 Rev. J...
AP304 and AP305 shaft and valve Stainless steel pH glass types Bulb general duties Flat in-line, self-cleaning Blue low temperature Reference system Ag/AgCl-3.5M KCl in gel matrix Reference junction Porous PTFE DS/AP300-EN Rev. I AP300 | Combination pH/Redox (ORP) sensors | IM/AP300 Rev. J...
Page 20
We reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document.