Appendix; Technical Data - THORLABS PM100USB Operation Manual

Optical power meter
Hide thumbs Also See for PM100USB:
Table of Contents

Advertisement

5 Appendix

5.1 Technical Data

General Data
Detector Compatibility
Display Type
Display Update Rate (max)
Display Format
Current Input (Photodiode Sensors)
Connector
Units
Measurement Ranges
Display Resolution
AD Converter
Accuracy
Input Bandwidth (Analog)
Wavelength Correction
Beam Diameter Setting
Voltage Input (Thermopile Sensors)
Connector
Units
Measurement Ranges
Display Resolution
AD Converter
Accuracy
Input Bandwidth (Analog)
Wavelength Correction
Beam Diameter Setting
© 2017 Thorlabs
Photodiode Sensors S100C Series
Thermal Sensors S3xxC / S4xxC Series
Pyroelectric Sensors EC100C/ES200C Series
Photodiodes (max. 5 mA)
Thermopiles (max. 1 V)
Pyros (max. 100 V)
External PC - Windows Application
Up to 300 Hz, depending on PC and settings
Numerical, Bar Graph, Trend Graph, Statistics, Histogramm
(Utility Software)
DB9F, left side
W, dBm, W/cm², A
6 decades; 50 nA - 5 mA
Ranges selectable in W or A, sensor depending
1 pA / responsivity value (A/W)
± 0.2 % f.s. (5µA - 5mA)
± 0.5 % f.s. (50nA)
DC - 100kHz, depending on sensor and settings
DB9F, left side
W, dBm, W/cm², V
4 decades; 1 mV - 1V
Ranges selectable in W or V, sensor dependent
1 µV / responsivity value (V/W)
± 0.5 % f.s. (10 mV - 1 V range)
± 1.0 % f.s. (1 mV range)
DC - 10Hz, depending on sensor and settings
Sensor depending; nm, (V/W)
16 bit
nm (A/W)
1/e²
16 bit
1/e²
5 Appendix