Pipe Wall; Path Length; Axial Length; Reynolds Correction - GE DigitalFlow GF868 Startup Manual

Ultrasonic flowmeter for flare gas
Hide thumbs Also See for DigitalFlow GF868:
Table of Contents

Advertisement

Chapter 2. Initial Setup

2.6.3 Pipe Wall

4. Use the numeric keys to enter the known thickness of the Pipe Wall. Press
If the pipe wall thickness is not available, look up the value in a table of standard pipe size data or use the Model
GF868's on-line Help Menu (see the Programming Manual for details).

2.6.4 Path Length

5. Press
= inch or
[F1]
(The meter will only accept values from 1/8 to 900 in.)
[ENT].
Note: The factory has calculated both the transducer signal path length
, based on the exact transducer configuration used for the application. These values are engraved on the
(L)
flowcell and/or are included in the documentation supplied with the meter.

2.6.5 Axial Length

6. Press
[F1]
= inch or
[F2] = feet
[ENT].
Note: The factory has calculated both the transducer signal path length
, based on the exact transducer configuration used for the application. These values are engraved on the
(L)
flowcell and/or are included in the documentation supplied with the meter.

2.6.6 Reynolds Correction

7. Press
to turn Reynolds Correction off, or
[F1]
Note: Reynolds Correction is a number based on the Kinematic Viscosity and flow rate of the fluid. It should be
enabled for most applications.
a. When you enable the Reynolds Correction Factor, you must also enter the Kinematic Viscosity of your gas, as
listed in Sound Speeds and Pipe Size Data. Use the numeric keys to enter a value, and press
8. Enter a value for the flow Calibration Factor and press
values from 0.5000 to 2.0000.)
28
to select the units. Then, enter the Path Length (P) of the ultrasonic signal. Press
[F2] = feet
to select the units.Then, enter the Axial Length (L) of the ultrasonic signal and press
to turn it on.
[F2]
. The default value is 1.00. (The meter will only accept
[ENT]
[ENT].
and the transducer signal axial length
(P)
and the transducer signal axial length
(P)
DigitalFlow™ GF868 Startup Guide (1 and 2-Channel)
.
[ENT]