Setting Up Ethernet Communications - GE DigitalFlow GF868 Programming Manual

Ultrasonic flowmeter for flare gas (1-channel)
Hide thumbs Also See for DigitalFlow GF868:
Table of Contents

Advertisement

Appendix C. Programming with PanaView
C.3.1 Setting up Ethernet Communications
If you have selected TCP/IP in step 6 on the previous page, the Setup Communications window appears similar to
Figure 19.
1. Type in the desired Name and Timeout (in milliseconds).
2. In the IP Address text box, enter the IP address. If the IP address is unknown, run the
utility. All units found will be identified by their MAC address and the assigned IP address. In the Port Number
box, enter 2101 (the default value).
3. Click
to complete data entry.
[OK]
IMPORTANT:
If you are using Ethernet communications, be sure the GF868 is set up with the default communications
parameters: 9600 baud, no parity, no handshaking, 8 data bits and 1 stop bit. The port number must
match the "Enable Raw TCP access using TCP Port" assignment shown under the
Server Settings" menu (
98
Figure 19: Setup Communications for TCP/IP
Configuration>Serial Ports>Port
Device Discovery
).
DigitalFlow™ GF868 Programming Manual (1-Channel)
software
"TCP
Device Discovery