Mjb8 - Linak LA20 User Manual

Linear actuators and electronics
Hide thumbs Also See for LA20:
Table of Contents

Advertisement

10. MJB8 (MEDLINE
CARELINE
®
Recommendations
• Always use locking mechanism and O-ring .
• Sockets that are not used must be fitted with blind plugs (Item P/N .: 0821008) to ensure IP-degree .
• When using the modular plug cable with an open end, the customer is responsible for maintaining the IP degree .
• When mounting, ensure that a screw torque limit of 1 Nm is not exceeded .
• 3
party products must correspond to all requirements stated in the MJB8 Interface description for 3
rd
malfunction .
• The Under Bed Light (Item P/N .: 0964135) must be mounted on the bed with metal screws in order to maintain ESD protection .
• HOT PLUGGING
Removing or adding any OpenBus™ cables are not allowed when the control box is powered by mains supply!
Follow the below procedure:
1 . Remove mains and wait 5 seconds
2 . Mount or dismount the required cables
If this procedure is NOT followed it may result in a damaged OpenBus
The risk of a damaged circuit increases if the accessory has a high start current .
• Before the final functional production test, it is important that the system is repowered .
This is to ensure, that all items have been detected on the OpenBus
• It is important to test the specified notifications in order to ensure that they work correctly before sending the system to the end user .
• The MJB8 is intended for mains operation only . The user must be informed that all MJB8 sensor notifications are disabled when the system is in
battery mode .
Warnings
• LINAK only takes responsibility for LINAK products, not 3
Please pay attention to the "Patient Environment" Clause 3 .79 – EN60601-1 3
It is important to inform the customer about this .
• The MJB8 is not able to detect defective 3
We recommend the end user to make a regular test procedure in order to prevent hazardous situations for the user and failures to the system .
• The MJB8 and the attached accessories (i .e . Out Of Bed or WET detection) are not intended as life-supporting or emergency equipment .
They are only intended to support notifications for communication and comfort purposes of people in need of care .
)
®
The MJB8 modular junction box is a central unit in the "Intelligent Care Bed", connecting various
intelligent accessories, such as the Out Of Bed and WET detection, and sending notifications via a range
of optional gateways to the user (i .e . caregiver)
party products .
rd
party products .
rd
driver circuit .
TM
.
TM
rd
edition . It must be subject to the Risk Analysis .
Page 181 of 192