ABB SwirlMaster FSS430 Commissioning Instructions page 34

Vortex & swirl flowmeter
Hide thumbs Also See for SwirlMaster FSS430:
Table of Contents

Advertisement

Operating mode /
Description
(order code)
Standard volume flow
Gas Std/Norm Vol. /
NG2
Gas Mass (reference
Gas mass flow, calculated using
density) / NG3
the reference density
Gas Mass (actual
Gas mass flow, calculated using
density) / NG3
the actual density
2) The highest priority of the device is to record the density via the analog input, as long as the analog input is activated as a density input. If the analog input is not
available as a density input, the system attempts to record the density via the HART input. If both the analog input and the HART input are deactivated as a density
input, the system uses the default density value.
3) The connection via the analog input or HART input is described in the chapter titled „Electrical connections" on page 21.
5) The highest priority of the device is to record the temperature via the analog input, as long as the analog input is activated as a temperature input. If the analog input is
not available as a temperature input, the system attempts to record the temperature via the HART input. If both the analog input and the HART input are deactivated
as a temperature input, the system uses the default density value.
7) If the selection "Gas linear." is set for the menu item Device Setup / Plant/Customized -> Gas Std. Mode, the compression factor is reset to 1.0. See also the chapter
titled "Special operating modes" in the operating instruction.
34 CI/FSS/FSV430/450-EN Rev. D | SwirlMaster FSS430, FSS450 VortexMaster FSV430, FSV450
Additional parameters required
Operating pressure
3) 5)
3) 5)
Operating temperature
Reference compression factor in the
normal condition
(AGA / SGERG only)
Operating compression factor
Reference pressure and reference
temperature in the normal condition
Reference density
Operating density
2) 3)
Parameter setting
Via analog input:
Input/Output / Field Input / Analog In Value ->
Pressure
Via HART input:
Input/Output / Field Input / Hart In Value ->
Pressure
Default setting for the pressure value:
Device Setup / Plant/Customized /
Compensation Setting -> Preset Pressure
With internal temperature sensor.
No information required, the measured value
from the temperature sensor is used.
Default setting for the temperature value:
Device Setup / Plant/Customized /
Compensation Setting -> Preset Int.Temp
7)
Adjustment via DTM/EDD
7)
Adjustment via DTM/EDD
Device Setup / Plant/Customized / Gas Ref.
Conditions
Device Setup / Plant/Customized / Gas Ref.
Conditions, as a selection for "Ref. Density "
Via analog input:
Input/Output / Field Input / Analog In Value ->
Density
Via HART input:
Input/Output / Field Input / Hart In Value ->
Density
Default setting for the density:
Device Setup / Plant/Customized /
Compensation Setting -> Preset Density