ABB SwirlMaster FSS430 Commissioning Instructions page 25

Vortex & swirl flowmeter
Hide thumbs Also See for SwirlMaster FSS430:
Table of Contents

Advertisement

Analog input 4 ... 20 mA
An external pressure transmitter, an external temperature
transmitter or a gas analyzer for a density signal can be
connected onto the analog input (4 ... 20 mA). The analog
input measures the current loop value.
— Temperature input for the return for power measurement
— Pressure input for pressure compensation for the flow
measurement of gases and steam
— Concentration of gas for the net metering of methane
(biogas)
— Density for calculating the mass flow
Current input
Terminals
Operating voltage
Input current
Equivalent resistance
1
P/N :
USE WIRING RATED
5ºC MIN. ABOVE MAX.
AMBIENT TEMPERATURE
TEST
PWR/COMM
+
EXT.
METER +
+
2
-
Fig. 25: Connection of transmitters at the analog input (example)
1 Remote transmitter
2 Power supply for the remote transmitter
3 Cable entry for the analog input 4 FSS400 / FSV400
5 Cable entry for the current output
6 Power supply for FSS400 / FSV400
ANALOG INPUT+ / ANALOG
INPUT-
16 ... 30 V DC
3.8 ... 20.5 mA
90 Ω
3
5
4
+
+
ANALOG INPUT
USE WIRING RATED
PWR / COMM.
5ºC MIN ABOVE MAX
AMBIENT TEMPERATURE
TEST
P/N:XXXXXXXXXXXX
EXT
METER+
DIGITAL
NAMUR-NO
NAMUR-YES
DIGITAL
OUTPUT+
OUTPUT–
1
4
2
3
SwirlMaster FSS430, FSS450 VortexMaster FSV430, FSV450 | CI/FSS/FSV430/450-EN Rev. D 25
HART 7 communication with remote transmitter
Because the device is designed with two-wire technology, a
pressure or temperature transmitter can also be connected to
the power supply lead (4 ... 20 mA).
The transmitter connected must run in HART burst mode. The
device supports remote transmitters with HART 7
communication.
+
1
3
P/N :
USE WIRING RATED
5ºC MIN. ABOVE MAX.
AMBIENT TEMPERATURE
TEST
PWR/COMM
EXT.
METER +
Fig. 26: Connection of transmitters with HART communication
(example)
6
1 Power supply for the remote transmitter
+
-
2 Power supply for FSS400 / FSV400
3 Remote transmitter 4 FSS400 / FSV400
5 Cable entry for the current output
G11772
-
4
+
+
ANALOG INPUT
USE WIRING RATED
PWR / COMM.
5ºC MIN ABOVE MAX
AMBIENT TEMPERATURE
P/N:XXXXXXXXXXXX
TEST
EXT
+
METER+
DIGITAL
NAMUR-NO
NAMUR-YES
DIGITAL
OUTPUT+
OUTPUT–
1
2
3
4
+
-
2
5
G11773

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents