ABB SwirlMaster FSS430 Commissioning Instructions page 31

Vortex & swirl flowmeter
Hide thumbs Also See for SwirlMaster FSS430:
Table of Contents

Advertisement

Easy Setup
Damping Qv
3.00s
Next
28. Use
to call up the edit mode.
29. Use
/
to set the damping for the volume flow.
30. Confirm the selection with
Easy Setup
Iout at Alarm
Low Alarm
Next
31. Use
to call up the edit mode.
32. Use
/
to select the alarm current.
33. Confirm the selection with
Easy Setup
Low Alarm
3.6000 mA
Next
34. Use
to call up the edit mode.
35. Use
/
to set the alarm current for "Low Alarm".
36. Confirm the selection with
Easy Setup
High Alarm
21,000 mA
Next
37. Use
to call up the edit mode.
38. Use
/
to set the alarm current for "High Alarm".
39. Confirm the selection with
Change from two to one column
Edit
.
Edit
.
Edit
.
Edit
.
SwirlMaster FSS430, FSS450 VortexMaster FSV430, FSV450 | CI/FSS/FSV430/450-EN Rev. D 31
Easy Setup
Auto Zero
3.6000 mA
Next
40. Use
to start automatic balancing of the zero point for
the system.
NOTE
Prior to starting the zero point balancing, make sure that:
— There is no flow through the sensor (close all valves,
shut-off devices, etc.)
— The sensor is completely filled with the medium to be
measured
Easy Setup
Low Flow Cutoff
0.00 %
Next
41. Use
to call up the edit mode.
42. Use
/
to set the desired value for the low flow cut-
off.
43. Confirm the selection with
Menu
Easy Setup
Exit
Select
Once all parameter have been set, the main menu appears
again. The most important parameters are now set.
44. Use
to switch to the process display.
OK