Download Print this page

Digital Data Output - Tektronix 067-1011-00 Instruction Manual

Digital sweep generator

Advertisement

067-1011-00 DSG
Step 5..
Check DIGITAL DATA OUTPUT
a)
Install a 7D01 Logic Analyzer/DF2 Display
Formatter Combination into a 7000 Series
oscilloscope and allow it to warm up.
b)
Connect the Logic Analyzer data acquisition probes
to the pins of the DIGITAL DATA OUTPUT connector
from the backside according to Table 4-3.
TABLE
4-3
7D01 HOOKUP ARRANGEMENT
DIGITAL DATA OUTPUT
(25 pin D Type Connector)
7D01 Input Probes
Connector Pin Number
Microclip Number
--------------------------------- -------------------------------
--------------------------------- -------------------------------
12
C (CLK signal-gray wire)
10
0 (LSB-black wire)
9
1 (brown wire)
8
2 (red)
7
3 (orange)
6
4 (yellow)
5
5 (green)
4
6 (blue)
3
7 (purple)
13
Gnd(white)
2
8 (black)*
1
9 (MSB-brown wire)*
13
Gnd*(white)
*2nd probe of microclips, for CH8-15 on 7D01.
NOTE:
Probe microclips will be attached to back side
of DIGITAL DATA OUTPUT connector.
c)
Set the Digital Sweep Generator CLOCK to INTernal
mode.
4 .. 7