Probe
mounting A
A
3.5
Mounting the probe
H
For use in pressure-free media only
To ensure overall accuracy, the thermostat must only be used in conjunction
with the pockets supplied by the factory.
The temperature probe (2) must be completely immersed in the medium. If this
is not the case, there will be appreciable deviations from the switching point.
With mountings U, UH and US, the temperature probe is secured in the pocket
by a clip (1).
3 Mounting
( 1 )
( 2 )
13