7.Put the load cell supporter module back, cover upper housing and tighten with the fixing
screws, then put the scale pan properly.
8.Proper setting is required(i.e. Baut rate, external device and print mode);
9. Wire connection between the external devices, please refer to 3.

6-5 Keypad Replacement

1. Remove the weighing pan.
3. Open the upper cover.
5.Remove the upper house and tear
off the keypad.
7.Please insert the keypad .
26
JADEVER SCALE CO., LTD
2. Put the scale upside down, loosen and remove
the 4 fixing screws.
4. Remove the keypad from the main board.
6.After the removal of residual glue,
attach the new keypad.
8. Put the upper housing on the scale
and test the keypad (refer to 4-6)
JCL Service Manual