Navigator 540 Transmitter - ABB Navigator 550 Operating Instructions Manual

Hydrazine analyzer, measurement made easy
Hide thumbs Also See for Navigator 550:
Table of Contents

Advertisement

Navigator 500
Hydrazine analyzer

2.2 Navigator 540 transmitter

The Navigator 540 transmitter is designed for continuous monitoring and control of power station boiler feed water / steam
condensate and must be used in conjunction with an associated ABB wet-section to measure levels of low level dissolved oxygen /
sodium / hydrazine. Wet-sections are parameter-specific.
Information from the wet-section is sent to the transmitter via a communication board, where the process reading is displayed on the
main page and can be displayed as a graph in the Chart View – refer to Section 6.7, page 30 for details of view options.
Diagnostic messages inform the user of the analyzer status and can be logged for review. The analyzer status can also be assessed
remotely using programmable alarms and current output diagnostic functions using optional Ethernet communications.
The transmitter has a multiple wet-section capability which enables it to control and display information from up to 4 wet-sections
(excludes multi-stream wet-sections). Section 3.3, page 13 shows an example of the multiple wet-section setup.
Mounting
bracket
Display
Keypad
Transmitter
door lock
Gland / cable entries
Fig. 2.2 Navigator 540 transmitter – main components
OI/AHM550–EN Rev. A
7

Advertisement

Table of Contents
loading

This manual is also suitable for:

Navigator 500 seriesNavigator 540

Table of Contents