Navigator 500
Hydrazine analyzer
7.1.4 Display
Menu
Display
Select
Exit
Menu
Language
Operator Templates
Page 1 (to 5) Template Page 1 template cannot be set, this defaults to the number of
Chart View
Channel 1 (to 4)
Source
Tag
Chart Duration
View/Log Enables
Continued on next page...
44
Used to select the display language, setup Operator page templates (1 to 5), enable diagnostic,
view and log functions, set the device's display brightness / contrast and set the time and date.
Comment
Selects the display language:
English / German / French / Italian / Spanish.
Enables a template to be assigned to a wet-section for display
purposes – refer to Section 6.5, page 28 for Operator Template
examples.
wet-sections fitted.
The chart can be configured to display the trend for 1, 2, 3 or 4
analog values. The engineering ranges for the process variable
values are configured in the Sensor Setup menu – see page 41.
Note. This menu is displayed only when Chart View is Enabled at the
Operator Functions / Chart View.
Selects the process variable signal to be displayed on the chart.
A 3-character, alphanumeric tag used to identify the parameter on
the chart.
Selects the chart duration from 1, 2, 4, 8, 12, 16, 20 and 24 hours.
Enables / disables the following View and Log functions:
View functions:
Diagnostics View / Signals View / Chart View / Alarm View / Analog
OP View – refer to Section 6.6, page 30 for examples of Operator
Pages in View mode.
Log functions:
Calibration Log / Alarm Log / Audit Log / Diagnostics Log – refer to
Section 6.7, page 31 for examples of Operator Pages in Log mode.
Default
English
Page 1