Hydrazine analyzer
...Range Change Setup
OI/AHM550–EN Rev. A
Analog output operation
Analogue output 2 is used for the range changing functionality and is
scaled depending on the range selected. The mA range is set
automatically to 4 to 20 mA.
Analog output 2
Range
Zero
1
0 % of mA range
2
0 % of mA range
3
0 % of mA range
Note. When range changing is enabled, the Analog Output 2
parameter option in the Input / Outputs level is not available –
see Section 7.1.5, page 46.
Relay operation
Relays 3 and 4 are used for the range changing functionality.
Connect Relay 3 – N/O to Relay 4 – COM
The range selection outputs can be determined as shown in the
following table:
Range
1
Relay 3 – N/C
2
Relay 4 – N/C
3
Relay 4 – N/O
SK1
RS485 SMART SENSOR
INTERFACE MODULE
ANALOGUE OUTPUT BOARD
CM40/0215 ISS.1
CM40/0235
SK4
SK5
SK5
RELAY
RELAY
OUT 3 OUT 4
5
6
SK2
I/O MODULE 2
I/O MODULE 3
I/O MODULE 3
1
2
3
4
5
+
+
N
L
COMMS
OUT 1 OUT 2
RELAY 1
RELAY 2
RELAY 3
RELAY 4
Relay 3
Relay 4
Note. When range changing is enabled, Relay 3 and 4 parameter
options in the Input / Outputs level are not available –
see Section 7.1.5, page 46.
Span
100 % of mA range
100 % of mA range
100 % of mA range
Voltage-free contact
Relay 3 – COM
Relay 3 – COM
Relay 3 – COM
43