Brief Procedures
Self Tests
Verify Internal Adjustment,
Self Compensation, and
Diagnostics
TDS6000B Series Specifications and Performance Verification Technical Reference
The Self Tests use internal routines to confirm basic functionality and proper
adjustment. No test equipment is required to do these test procedures.
The Functional Tests utilize the probe-compensation output at the front panel as
a test-signal source for further verifying that the oscilloscope functions properly.
A BNC cable and a BNC-to-SMA adaptor or a TCA-SMA adapter are required
to do these test procedures.
This procedure uses internal routines to verify that the oscilloscope functions and
was adjusted properly. No test equipment or hookups are required.
Equipment
None
required
Prerequisites
Power on the oscilloscope and allow a 20 minute warm-up before doing
this procedure.
1. Verify that internal diagnostics pass: Do the following substeps to verify
passing of internal diagnostics.
a. Display the System diagnostics menu:
H If the oscilloscope is in toolbar mode, click the MENU button to put
the oscilloscope into menu bar mode.
H Select Instrument Diagnostics from the Utilities menu and. This
displays the diagnostics control window.
b. Run the System Diagnostics:
H First disconnect any input signals from all four channels.
H Click the Run button in the diagnostics control window.
c. Wait: The internal diagnostics do an exhaustive verification of proper
oscilloscope function. This verification may take several minutes. When
the verification is finished, the resulting status will appear in the
diagnostics control window.
3- 1