Appendix A) Troubleshooting Tips

Make sure the SCPI server has been started on the instrument.
'Reset LCA System' and 'Reset Remote Interface' from the LCA Server as below if the
expected behavior is not observed.
Test individual SCPI commands using the Keysight Connection Expert; this is a good
approach to verify command syntax. If sending a command to set an instrument
parameter (i.e. ':PARameter:WAVelength 1310') immediately query the set value to see
if the LCA recognized the command (i.e. ':PARameter:WAVelength?'). Additionally, for
every command which is tested in the Keysight Connection Expert, send the command to
query for system errors (i.e. ':SYST:ERR?'). This will help identify if there is a portion of the
command which is not recognized by the instrument.