Smeg WD3060 User Manual page 29

Washer disinfector
Hide thumbs Also See for WD3060:
Table of Contents

Advertisement

ALARM ID
DISPLAY MESSAGE
"Temperature
ID77
> 45"
"Recovery
ID78
failed"
"Program
ID79
not congruent"
"Warm water
ID80
flowmeter failure"
"Demi water
ID81
flowmeter failure"
"System failure
ID82
sol 1"
"Overtemperature
ID84
TA1"
"Archive
ID91
full"
"Change
ID92
filter"
ID93
"Service"
"Temperature
ID94
floating"
WD3060 - WD5090
MEANING AND SUGGESTED USER ACTION
Intake water temperature over 45°C; prewash temperature must
be below 45°C.
.Wait for the appliance to cool before starting a new cycle.
.Check the cold water temperature
.Verify the correct setting of the water supply.
Microprocessor error.
Call the Technical Assistance Service if the fault persists.
This alarm is triggered if a program is created using the
WDTRACE software using settings not compatible for the proper
performance of the cycle.
.Reconsider the program created phase by phase
Warm Water, intake control system failure.
.Check the water supply settings.
Demi Water, intake control system failure.
.Check the water supply settings.
Door locking mechanism failure.
1. Make sure the door is closed properly before starting a cycle.
2. Do not force the door open with a cycle in progress; always
use the buttons provided on the appliance to stop a cycle and
open the door.
3. Check there is nothing between the door and the appliance's
chamber preventing the door from closing properly.
4. Follow the DEFAULT PROCEDURE described above.
Anomalia nella chiusura porta della vasca.
Dryer temperature reading higher than target temperature.
. Follow the DEFAULT PROCEDURE described above.
Internal memory full.
.Memory space must be cleared to allow use of the appliance to
continue: this can be done by connecting to the appliance by
means of the RS-232 serial port and using the WDTRACE
software.
The absolute filter in the drying system has exceeded the set
number of working hours. Contact the After-Sales Service to
have the filter changed. The filter is a consumable and is not
covered by the warranty.
The appliance has exceeded the set number of operating hours
since installation or since its last service: contact the After-Sales
Service for routine maintenance.
Temperature below TARGET temperature during extension
phase.
. Follow the DEFAULT PROCEDURE described above.
USER MANUAL
29