Smeg WD3060 User Manual page 27

Washer disinfector
Hide thumbs Also See for WD3060:
Table of Contents

Advertisement

ALARM ID
DISPLAY MESSAGE
"Washing chamber drainage
ID29
failed"
ID30
"Safety chamber level exceeded"
"Safety level
ID31
Safety level failed"
"Sump full
ID32
Turn the taps off"
"Lack of water
ID33
in the Condenser"
"Condenser Drainage
ID34
failed"
"Not reached Target
ID37
Cooled Drainage"
"Cooled Drainage
ID38
failed"
"Detergent 1
ID41
Inflow failed"
ID42,
"Detergent 2, 3, 4, or 5
ID43,
Inflow failed"
ID44, ID45
"Pump tube 1
ID46
clogged"
WD3060 - WD5090
MEANING AND SUGGESTED USER ACTION
Chamber does not empty. No drainage.
.Check the connection to the water drain,
especially that the height of the drain connections complies with
the specified requirements and that there are no restrictions in
the drain hoses.
During the working cycle, the chamber water level exceeds the
safety level.
Check the water supply to the appliance:
1. Intake pressure.
2. That the connections are correctly set, as specified.
.The same checks as for ID30
Water standing in washing chamber with appliance in standby.
.The same checks as for ID30
Steam condenser water intake anomaly.
Check the water supply:
1. Intake water pressure
2. That the connections are correct, as specified in this manual.
3. Blockages or restrictions in hoses.
4. Check intake water hardness (hard water contribute to steam
condenser nozzles obstruction).
Steam condenser water drainage failure.
Check that the drain connections are correct:
1. Height and position of the drain connection
2. That the connections are correct, as specified in this manual.
3. Blockages or restrictions in hoses.
Drainage problems when using a mixture of water. The use of a
mixture of water is an option adopted in special cases to cool the
water discharged into the drain.
Check:
1. The intake water temperature; if the water is not cold enough
(recommended T<25°C) there may be difficult in achieving the
target temperature.
2. Problems with drain; check that the hoses and drain
connections meet the specified requirements.
.The same checks as for ID37
. Check the detergent presence.
. Check that the detergent intake nozzle is correctly positioned in
the tank.
. Check for detergent leaks (pools of detergent around the
device).
.The same checks as ID41, for the detergents 2, 3, 4, 5.
.The same checks as for ID41
USER MANUAL
27

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wd5090

Table of Contents