ALARM ID
DISPLAY MESSAGE
ID5
"Probe TL1 disconnected"
ID7
"Probe TA1 disconnected"
ID10
"Probe TCL disconnected"
ID11
"Lack of cold water"
ID12
"Lack of warm water"
"Lack of demi water"
ID13
ID17
"Cold water load time exceeded"
"Warm water load time
ID18
exceeded"
ID19
"Demi water load time exceeded"
ID20
"Water load system failure"
ID22
"Cold water flowmeter failure"
ID23
"Insufficient water in chamber"
ID24
"NO water in chamber"
ID25
"No Press in Hyd Sys Foam(?)"
WD3060 - WD5090
MEANING AND SUGGESTED USER ACTION
Working temperature probe "TL1" is generating an abnormal
signal (probe "open").
.follow the DEFAULT PROCEDURE described above.
Drying temperature probe TA1 is generating an abnormal signal
(probe "open").
.follow the DEFAULT PROCEDURE described above.
Control temperature probe TLC is generating an abnormal signal
(probe "open").
.follow the DEFAULT PROCEDURE described above.
No cold water during filling phases.
.Check the water supply.
.Check that the intake tap is open.
.Check the water supply pressure.
.Check that the filler hoses are correctly positioned.
The same as ID11 for the warm water.
.Check the correct setting of the parameters of the device
(effective presence/absence of warm water).
The same as ID11 for the demi water.
.Check the correct setting of the parameters of the device
(effective presence/absence of demi water).
Cold water filling time not correct.
.The same checks as for ID11
Warm water filling time not correct.
.The same checks as for ID12
Demineralized water filling time not correct.
.The same checks as for ID13
.The same checks as for ID11, 12, 13
.The same checks as for ID11
Chamber water level too low.
.The same checks as for ID11, 12, 13
. Verify that there are no leaks from the machine, highlighted by
the presence of water near the device.
.The same checks as for ID23.
Anomaly related to the circulating pump "ML". Circulating pump
pressure too low. There may be foam in the chamber.
.Check the type of detergent used.