ALARM ID
DISPLAY MESSAGE
ID47,
"Pump 2, 3, 4 or 5 tube
ID48,
clogged"
ID49, ID50
ID52
"Door 1 electrically open"
"Door 1
ID54
mechanically open"
"Doorlock 1
ID56
failure"
"Drying 1
ID58
failure"
"Drying 1
ID60
Safety Switch on"
"Heating
ID62
Thermostat on"
"Cooling failed
ID67
WARNING: high Temp"
"Tank 1
ID68
empty"
ID69,
"Tank 2, 3, 4 or 5
ID70,
empty"
ID71, ID72
ID73
"Archive error"
"Lack of
ID75
salt"
WD3060 - WD5090
MEANING AND SUGGESTED USER ACTION
.The same checks as ID41, for the detergents 2, 3, 4, 5.
Door locking mechanism failure.
1. Make sure the door is closed properly before starting a cycle.
2. Do not force the door open with a cycle in progress; always
use the buttons provided on the appliance to stop a cycle and
open the door.
3. Check there is nothing between the door and the appliance's
chamber preventing the door from closing properly.
4. Follow the DEFAULT PROCEDURE described above.
Door opening detected with cycle in progress. Door interlock
microswitch malfunction.
.The same checks as for ID52.
.The same checks as for ID52.
Drying heating system failure.
.follow the DEFAULT PROCEDURE described above.
.Check the condition of the safety thermostat (push the second
red plunger on the upper wall of the detergent cabinet to reset
the thermostat).
Drying heating system failure.
.Check the condition of the safety thermostat (push the second
red plunger on the upper wall of the detergent cabinet to reset
the thermostat).
Chamber Heating elements safety device intervened.
.Check the condition of the safety thermostat (push the first red
plunger on the upper wall of the detergent cabinet to reset the
thermostat).
Dryer motor "cooling" malfunction. "Cooling" phase is included
at the end of the drying phase to bring the load processed and
the heating elements to a safe temperature.
.follow the DEFAULT PROCEDURE described above.
.Check that there is detergent in the jerry can P1 or that the level
sensor is operating correctly.
.Check that there is detergent in the jerry can or that the level
sensor is operating correctly.
Internal memory data storage error.
.follow the DEFAULT PROCEDURE described above.
.Add salt to the softener salt tank inside the chamber, accessible
with the door open.