DMM Option-Specification
24XSB12467B Options Service
Table 4-1 (cont)
Characteristics
Response Time
Auto Range
I
Less than 2 second^.^
I
Performance Requirements
Manual Range
Reading Rate
I
Approximately 3 per ~ e c o n d . ~
AMPS
I
Less than 1 ~ e c o n d . ~
DC Accuracy
I
AC Accuracy
+18"C to +28"C
-15°C to +18"C and +28"C to +55"C
20 Hz to 10 kHz sinusoidal waveform.
+
(0.6% of reading
+
0.1
O / O
of full scale).
+
(0.7% of reading
+
0.1 5% of full ~ c a l e ) . ~
+18"C to +28"C
1
k
(0.6% of reading
+
0.1
O / O
of full scale).
I
-15°C to +18"C and +28"C to +55"C
1
+
(0.7% of reading
+
0.1 5% of full ~ c a l e ) . ~
I
Response Time
Manual Range
I
Less than 1 ~ e c o n d . ~
I
Auto Range
I
Less than 2 second^.^
I
lnput Resistance by Range
100 pA
I
Approximately 1.0 kCLa
I
1 mA
I
Approximately 100.0
f l a
I
1 A (1 000 mA)
I
Approximately 0.5 C L a
I
10 mA
100 mA
Maximum Input Current
I
1 A.a
I
Approximately 10.5
f l a
Approximately 1.5 C a
Resolution
I
1 part in 10,000 full scale.a
CONTINUITY
TEMPERATURE
Response Time
Threshold Resistance
- - -
Accuracy
Approximately 0.1 ~ e c o n d . ~
10 Q k 1
aa
+
1 8" C to
+
28" C Ambient Temperature
-15°C to +18"C and +28"C to +55"C Ambient
Temperature
Resolution
I
0.1 "C or 0.1 " F . ~
+
(2% of reading
+
1.5" C).a
+(2% of reading
+
2.0°C).a
- -
Probe Tip Measurement Range
aPerformance Requirement not checked in manual.
-62°C to +230°C in one range.a