ECG Monitorinn
Press the ECG SIZE up-down arrows until the desired waveform size is displayed.
Adjust systole beeper tone volume to suitable level.
Set the Heart Rate Alarm by pressing the ALARM SET button. (Refer to the next
page for instructions on changing preset settings.) When the alarm is activated,
the line crossing out the bell-shaped character will disappear.
Press ALARM ON-OFF switch. This will activate heart rate alarms.
When an alarm occurs the recorder will auto-
matically run for 15 seconds, the bell-shaped
symbol will flash (the heart symbol freezes for
easier identification), and the audible alarm tone
will sound. Deactivating the alarms turns off the
tone and the flashing bell and reactivates the
heart symbol to flash with a detected R-wave.
SETTING ALARMS
Heart rate alarms have been preset at 30 (low) and 150 (high)
To change the lower or upper alarm set points:
1. Push ALARM SET button.
Low alarm limit value flashes.
To change value, push
A
to raise,
push
V
to lower
2. Push ALARM SET button again.
High alarm limit value flashes.
To change value, push
A
to raise,
push
V
to lower
3. Push ALARM SET button again to return to alarm monitoring mode.
RECORDER
The Recorder will document the ECG trace with a 6 second delay at all times.
OPERATION
To start the recorder, press the RECORDER
ON-OFF button beside the monitor. The recorder
will run continuously until the button is pressed
again. To stop the recorder, press the
RECORDER ON-OFF button again.