SPECIFICATIONS
Size
Weight
Power
Warranty
Design
Standards
Patient Safety
Environmental
General Information
GENERAL
10.7 cm high x 33.5 cm wide x 31 cm long (4.2 in. x 13.2 in x 12.2 in).
5.9 kg. (1 3 Ibs.) with PD 4400 multi-function cable; 6.8 kg (1 5 Ibs.) with paddles
Sealed lead acid battery
-
2.5V/cell, 5 cells -wired in series.
In North America: 1 year, including use of a loaner.
Outside North America: consult ZOLL authorized representative.
Meets or exceeds all AAMI specifications for defibrillators.
Meets or exceeds UL 544, IEC 601 and CSA standards for medical equipment and
MIL-STD-81 OD, Method 51 4.3, Figure 51 4.3-35 (Cat 6), Swept Sine, Random
Vibration, Universal Helicopter, MIL-STDBIOD, Method 514.3 (Cat. 3).
All patient connections are isolated.
Temperature:
0°C to 55°C (operating).
-40°C to 75°C (storage and shipping).
Humidity:
5% to 95% relative humidity.