13.7 Start screen

Turning the start screen on and off. When the device is switched on,
the start screen appears with your company logo. The company logo
can be uploaded to the device via the PC measurement data manage-
ment.

13.8 PIN

You can protect your device against unauthorized access by using
your personal 4-digit PIN.

13.9 Calorific value

If activated, negative losses (qA) and degrees of efficiency (ETA) exceeding 100% are con-
sidered in the measurement. This function should always be activated for condensing boiler
systems, so that measurement results are valid. This setting has an effect on the measure-
ments in the Quick menu.

13.10 Extended fuel list

The fuel list with the combustible materials light fuel oil, natural gas, propane, heavy
fuel oil and pellets is extended by the following combustible materials:
wood, brown coal, stone coal, stone coal briquettes, stone coal coke, anthracite coal,
biogas, butane, city gas, coke oven gas

13.11 Federal Immission Control Ordinance (BImSchV) average value

The 30 second average measurement according to BImSchV during the exhaust measure-
ment can be switched on or off.
13.12 NO
factor
x
This function only applies to devices with NO meter flume. It allows considering the per-
centage of NO
from NO
for the calculation of NO
. For example, if the NO
percentage
2
x
x
2
was determined with a 6% of the NO percentage by means of a measurement, the meas-
ured NO value must be multiplied with 1.06, to obtain NO
. For this purpose, the NO
fac-
x
x
tor must be set to 1.06.
29
5695061
Edition 01 – April 2016
en