Operating Instructions Dräger FG7000
The test pressure can be adjusted between 20 and 25,000 mbar for
tightness tests, the stabilization time and the measurement time be-
tween 5 minutes and 6 hours. Optionally available high pressure sen-
sors must be used for pressures exceeding 150 mbar.
The input function begins with the selection buttons behind the values.
"Next" starts the measurement program. During the measurement,
">>/<<" can be used at any time to switch between table and diagram
display.
Pressure built-up: In this phase, the measuring device waits for the test
pressure to build up. "Done" confirms that the test pressure has been
reached.
Stabilization phase: The measuring device waits until the specified sta-
bilization time lapses and then automatically switches to measurement.
The stabilization phase can be ended manually by using "Next".
Measurement: The pressure pattern as well as the starting, end and
differential pressure are recorded during the measurement period. The
selected measurement period can be shortened using "Quit".
Done: After the measurement, the recorded results are available.
21
5695061
Edition 01 – April 2016
en