ADAM-5018 7-ch Thermocouple Input Module
The ADAM-5018 is a 16-bit, 7-channel thermocouple input module that features pro-
grammable input ranges on all channels. It accepts millivolt inputs (±15 mV , ±50 mV ,
±100 mV , ±500 mV), voltage inputs (±1 V , ±2.5 V), current input (±20 mA, requires 125
ohms resistor) and thermocouple input (J, K, T , R, S, E, B).
The module forwards the data to the host computer in engineering units (mV , V , mA or
temperature °C). An external CJC on the plug-in terminal is designed for accurate tem-
perature measurement.
ADAM-5018
Application Wiring
ADAM-5000 Series User Manual
Figure 1.21 ADAM-5018 Module Front View
Figure 1.22 Thermocouple Input