5.
5.5
Replacing the Main Battery Pack
Periodically you must replace the main battery pack.
Figure 5-5. Main Battery Pack
The device provides a battery cable with two pairs of battery connectors, enabling you to hot-swap
the battery pack in a non-hazardous location. You attach the new battery pack to the free
connector (leaving the old one connected), and then disconnect the old battery pack.
DANGER
EXPLOSION HAZARD: Ensure the area in which you perform this operation is non-hazardous.
Performing this operation in a hazardous area could result in an explosion.
Service and Troubleshooting
Close the enclosure.
Emerson FB2200 Flow Computer Instruction Manual
D301784X012
March 2019
95