Reference   S Ection:   I Nitialization   M Ode   ( Init); Input   C Onfiguration   ( Init   >   I Npt); Thermocouple   I Nput   T Ype   ( Init   >   I Npt   >   T - Omega PLATINUM Series User Manual

Temperature, process and strain meters
Hide thumbs Also See for PLATINUM Series:
Table of Contents

Advertisement

PLATINUM   S eries   M eters   U ser's   G uide  
 
 
6. Reference   S ection:   I nitialization   M ode   ( INIt)  
Use   I nitialization   M ode   t o   s et   t he   f ollowing   p arameters   a nd   p erform   t he   f ollowing   f unctions:  
 
6.  
Reference   S ection:   I nitialization   M ode   ( INIt)   . .............................................................................   2 2  
6.1
 
Input   C onfiguration   ( INIt   >   I NPt)  ................................................................................................   2 2  
6.2
 
TARE   ( INIt   >   t ARE)   . .....................................................................................................................   2 9  
6.3
 
LINR   ( INIt   >   L INR)   . ......................................................................................................................   2 9  
6.4
 
6.5
 
Excitation   V oltage   ( INIt   >   E CtN)   . ................................................................................................   3 4  
6.6
 
Communication   ( INIt   >   C oMM)   ..................................................................................................   3 4  
6.7
 
Safety   F eatures   ( INIt   >   S Fty)   . .....................................................................................................   3 8  
6.8
 
Manual   T emperature   C alibration   ( INIt   >   t .CAL)   . ........................................................................   4 0  
6.9
 
6.10
6.11
  Display   F irmware   R evision   N umber   ( INIt   >   V ER.N)   . ...................................................................   4 1  
6.12
  Update   F irmware   R evision   ( INIt   >   V ER.U)  ..................................................................................   4 2  
6.13
  Reset   t o   F actory   D efault   P arameters   ( INIt   >   F .dFt)  ....................................................................   4 2  
6.14
  Password-­-Protect   I nitialization   M ode   A ccess   ( INIt   >   I .Pwd)   . .....................................................   4 2  
6.15
  Password-­-Protect   P rogramming   M ode   A ccess   ( INIt   >   P .Pwd)   . ..................................................   4 2  
6.1
Input   C onfiguration   ( INIt   >   I NPt)  
Select   t he   I nput   p arameter   ( INPt)   t o   c onfigure   t he   i nput.  
 
Navigate   t o   t he   c orrect   s etting.   S ettings   i nclude   t he   f ollowing:  
 
t.C.  
-   Thermocouple   T emperature   S ensor   ( entry   p oint)  
Rtd  
-   Resistance   T emperature   D etector   ( RTD)  
tHRM   -   Thermistor   T emperature   S ensor  
PRoC   -   Process   V oltage   o r   C urrent   I nput  
Select   t he   i ndicated   s etting.  
 
6.1.1
Thermocouple   I nput   T ype   ( INIt   >   I NPt   >   t .C.)  
Select   T hermocouple   ( t.C.)   a s   t he   i nput   t ype   ( factory   d efault).   T hen   s pecify   a   s pecific   t ype   o f  
 
thermocouple   o r   t he   l ast   s elected   t ype   w ill   b e   u sed.  
Navigate   t o   t he   i nstalled   t hermocouple   t ype.   S upported   t ypes   a re   a s   f ollows:  
 
k  
-   Type   K   ( factory   d efault)  
J  
-   Type   J  
t  
-   Type   T  
E  
-   Type   E  
N  
-   Type   N  
R  
-   Type   R  
S  
-   Type   S  
b  
-   Type   B  
C  
-   Type   C  
Select   t he   i ndicated   t ype.  
 
 
Reference   S ection:   I nitialization   M ode   ( INIt)  