PLATINUM   S eries   M eters   U ser's   G uide  
 
 
6.2
TARE   ( INIt   >   t ARE)  
Select   T ARE   ( tARE)   t o   c onfigure   t he   t are   o ption.  
 
Navigate   t o   t he   d esired   s etting.   S ettings   i nclude   t he   f ollowing:  
 
dSbL   –   No   T ARE   o ption   i s   e nabled  
ENbL   –   TARE   o ption   m ay   b e   a ctivated   f rom   R UN   m enu  
RMt   –   TARE   o ption   m ay   b e   a ctivated   f rom   R UN   m enu   o r   b y   d igital   i nput  
 
The   T are   o ption   i s   o nly   a vailable   f or   p rocess   i nputs.   I f   e nabled   ( ENbL   o r   R Mt)   t he   O PERATING  
menu   i s   e xpanded   t o   i nclude   a   T ARE   o ption   t hat   a llows   z eroing   t he   r eading   v alue   w hen   t he  
unit   i s   i n   t he   R UN   m ode.  
If   t he   r emote   o ption   i s   s elected   t he   r eading   m ay   b e   z eroed   u sing   e ither   t he   p anel   b utton   o r  
the   d igital   i nput   w hile   i n   t he   R UN   m ode.  
When   a ctivated,   t he   T ARE   p rocess   a pplies   a n   o ffset   a djustment   t o   t he   c urrent   r eading   s uch  
that   t he   a djusted   r eading   w ill   b e   z ero.  
Select   t he   i ndicated   s etting.  
 
6.3
LINR   ( INIt   >   L INR)  
Select   1 0   P oint   L inearization   o ption.   T he   1 0   p oint   l inearization   a llows   e ntering   u p   t o   1 0  
 
Reading/Input   v alue   p airs   u sed   t o   i nternally   c alculate   1 0   g ain/offset   p arameters.   T he   L INR  
option   i s   o nly   a vailable   o n   P rocess   i nputs.  
Select   t he   N umber   o f   P oints   ( N.PNt  
 
 
The   N .Pts   o ption   d etermines   h ow   m any   o f   t he   R eading/Input   p airs   a re   p rocessed.   T o   d isable  
the   L inearization   f unction   s et   t he   N umber   o f   p oints   t o   0 .  
Select   t he   i ndicated   s etting.  
 
Choose   e ither   m anual   o r   l ive   i nputs.   T he   d efaults   f or   e ach   i nput   p air   1 00.0   R eading   a nd   1 00.0  
 
Input.   S caling   m ethods   i nclude   t he   f ollowing:  
MANL   –   User   m anually   e nters   b oth   p arameters  
LIVE   –   User   m anually   e nters   t he   d isplay   v alues   ( RD.n)   b ut   t he   s ignal   l evel   i s   r ead  
from   t he   i nput   s et   t he   i nput   v alues   ( IN.n)  
Using   e ither   M ANL   o r   L IVE   l inearization   G ain/Offset   p airs   a re   c alculated   a s:  
Gain   =   ( Rd.n   –   R d.n-­‐1)   /   ( IN.n   –   I N.n-­‐1)  
Offset   =   R d.n-­‐1   –   ( Gain   *   I N.n-­‐1)  
During   p rocessing   t he   i nput   l evel   i s   c ompared   t o   t he   s et   o f   i nput   l evels   s tored   t o   d etermine  
which   g ain/offset   p air   t o   a pply.  
Select   t he   l inearization   m ethod   t o   b e   u sed.  
 
   
Navigate   t o   t he   d esired   s caling   p arameter.   O ptions   i nclude   t he   f ollowing:  
 
Rd.n   –   Reading   l ow   v alue   c orresponding   t o   I N.1   s ignal  
IN.n   –   Input   s ignal   c orresponding   t o   R D.1    
In   M anual   M ode,   I N.n   i s   e ntered   m anually.   I n   L ive   M ode,   t he   I N.n   o ption   d isplays   t he   c urrent  
input   s ignal.  
Select   t he   s caling   p arameter   t o   c hange.  
 
 
Reference   S ection:   I nitialization   M ode   ( INIt)  