Manual   T Emperature   C Alibration   ( Init   >   T .Cal) - Omega PLATINUM Series User Manual

Temperature, process and strain meters
Hide thumbs Also See for PLATINUM Series:
Table of Contents

Advertisement

PLATINUM   S eries   M eters   U ser's   G uide  
 
 
6.8
Manual   T emperature   C alibration   ( INIt   >   t .CAL)  
Select   t he   M anual   T emperature   C alibration   ( t.CAL)   s ubmenu.   T his   p arameter   a llows   y ou   t o  
 
manually   a djust   t he   t hermocouple,   R TD,   o r   T hermistor   c alibration   c urves   p rovided   w ith   t he  
unit.     O nce   a   c urve   h as   b een   m anually   a djusted,   t his   s etting   c an   b e   s et   t o   N oNE   t o   d isable   t he  
manual   a djustment   ( resetting   t o   f actory   d efaults   w ill   r emove   a ny   m anually   a djustment  
factors).        
Navigate   t o   t he   d esired   s etting.   S ettings   i nclude:  
 
NoNE   –   No   m anual   c alibration   ( factory   d efault)  
1.PNt   –   Manually   c reate   a   1 -­‐point   c alibration  
2.PNt   –   Manually   c reate   a   2 -­‐point   c alibration  
ICE.P   –   Manually   c reate   a   1 -­‐point   c alibration   a t   0
Select   t he   i ndicated   o ption.  
 
6.8.1
No   M anual   T emperature   C alibration   A djustment   ( INIt   >   t .CAL   >  
NoNE)  
Select   N oNE   t o   u se   t he   s tandard   t emperature   s ensor   c alibration   c urves.   T his   m ode   w ill   b e  
 
used   b y   m ost   u sers.  
6.8.2
Manual   T emperature   C alibration   O ffset   A djustment   ( INIt   >   t .CAL  
>   1 .PNt)  
Select   1 .PNt   t o   m anually   a djust   t he   o ffset   o f   t he   c alibration   c urve   b ase   o n   t he   c urrent  
 
reading.  
Set   t he   M anual   T hermocouple   C alibration   O ffset   v alue   i n   d egrees.  
 
Confirm   t he   O ffset   v alue   a nd   p air   i t   w ith   t he   c urrent   i nput   r eading.  
 
6.8.3
Manual   T emperature   C alibration   O ffset   a nd   S lope   A djustment  
(INIt   >   t .CAL   >   2 .PNt)  
Select   2 .PNt   t o   u se   2   p oints   t o   m anually   a djust   b oth   t he   o ffset   a nd   s lope   o f   t he   c alibration  
 
curve.  
Navigate   t o   t he   d esired   s etting.   S ettings   i nclude   t he   f ollowing:  
 
R.Lo   –   Set   l ow   p oint   i n   d egrees,   d efault   =   0 ,   a nd   a ssociate   w ith   i nput   r eading  
R.HI   –   Set   h igh   p oint   i n   d egrees,   d efault   =   9 99.9,   a nd   a ssociate   w ith   i nput   r eading  
Select   t he   i ndicated   s etting.  
 
Set   t he   T emperature   f or   R .Lo   o r   R .HI.  
 
Confirm   t he   v alue   a nd   p air   i t   w ith   t he   c urrent   i nput   r eading.  
 
 
 
 
 
Reference   S ection:   I nitialization   M ode   ( INIt)  