2.4 Commands
LCME?
Example:
PARI(?) {z}
Example:
TOKN(?) {z}
Example:
SIM984 Isolation Amplifier
Command Error
Query the last command error code. A query of LCME? always clears
the error code, so a subsequent LCME? will return 0. Valid codes are:
Value
Definition
No execution error since last LCME?
0
1
Illegal command
2
Undefined command
3
Illegal query
4
Illegal set
5
Missing parameter(s)
6
Extra parameter(s)
7
Null parameter(s)
8
Parameter buffer overflow
9
Bad floating-point
10
Bad integer
11
Bad integer token
12
Bad token value
13
Bad hex block
14
Unknown token
*IDN
LCME?
4
The error (4, "Illegal set") is due to the missing "?".
Parity
Set (query) parity {to z = (NONE 0, ODD 1, EVEN 2, MARK 3, SPACE 4)}.
After power-on, modules default to PARI NONE.
PARI EVEN
Token Mode
Set (query) the Token Query mode {to z=(OFF 0, ON 1)}.
If TOKN ON is set, then queries to the SIM module that return to-
kens will return the text keyword; otherwise they return the decimal
integer value.
Thus, the only possible responses to the TOKN? query are ON and 0.
On reset, TOKN is set to OFF.
TOKN OFF