81.
Operation | Replacing the Optical Filter
Replacing the Optical Filter
▪ Be careful not to touch the optical filter.
▪ Optical filter replacement should be done in a clean area. Ensure the
filter stays free from dust or dirt particles, and that dust or dirt particles
do not enter the camera when replacing the filter.
To replace the optical filter:
1. Remove the lens from the camera (see
2. Assemble the filter replacement tool by screwing the tool head (B)
onto the handle (A).
3. Insert the tool into the camera and turn counterclockwise (C) to
remove the optical filter which is currently in the camera.
4. Place the removed filter into the optical filter box.
5. Point the camera up and gently insert the new filter into the camera.
6. Using the tool, turn the optical filter clockwise until tight.
7. Reconnect the lens to the camera.
MENU
SAFETY
OVERVIEW
CAUTION
Replacing the
Lens).
OPERATION
TECHNICAL
SPECIFICATIONS
A
MAINTENANCE
TROUBLESHOOTING
www.
B
C
TECHNICAL
SUPPORT
EyeCGas® 2.0 | User Manual Rev 01 | Doc P/N: UM-8G9A0000B